Smart environmental monitoring solutions

Talk to our sales team. Call +61 2 9622 6327

Request a quote
Back to products

JCS-100

Sample Gas Cooler (Compressor)

  • For extractive gas analysis
  • For emission and process monitoring
  • Continuous drying of sample gas to a precise low and constant outlet dew point
  • Minimizes water vapor  cross sensitivities and volumetric errors

Have questions? Call +61 2 9622 6327

Features

  • Proven and reliable technology
  • Various heat exchanger materials
  • One or two independent gas paths
  • Digital temperature indication
  • Integrated condensate pump
  • Status contact with acoustic feedback
  • Hot gas bypass compressor technology
  • Temperature sensor monitoring
  • High current volt free status relay outputs (direct pump switching)
  • Long life fan with ball bearing
  • Wall mount, optional free stand
  • Fully encapsulated temperature sensor
  • Small footprint
  • Stainless steel housing
  • Light weight


Benefits

  • High flow rates and water vapour dew points possible
  • Easy to change heat exchangers
  • Continuous condensate removal
  • Low maintenance operation
  • For extractive gas analysis
  • For emission and process monitoring
  • Continuous drying of sample gas to a precise low and constant outlet dew point
  • Minimises water vapour cross sensitivities and volumetric errors
Descriptionsample gas compressor coolers
Cooling principlecompressor cooling
Number of heat exchangers / Gas paths1 or 2
Material heat exchangerscoated aluminum, PVDF, Duran glass, SS316

Operations

Max. gas flow rate” Nuhr250
Max. gas inlet temperature* f°Cl140
Max. gas inlet dew point* °CI80
Gas outlet dew point3° or 5°C
Dew point stability
(for constant inlet conditions)
+0.2 K
Ambient temperature5°to 45 °C
Cooling capacity160 W
Temperature alarm limits<0/> +10°C
Operating pressure with condensate pump0.2 to 2.2 bara
Max. operating pressure
without condensate pump bara
2.5
Ready for operation< 20 min.
Pressure drop at 100 NU/hrPressure drop at 100 NU/hr

Constructions

Dimensions over all (W x H x D)308 x312 x375 mm
Installationstand alone or wal mounung
Mounting positionhorizontal
Weight17 kg
Housing / Colourstainlesssteel
Dead volume per gas path67 ml
Connection sample gas and condensate
outlet with condensate pump
PVDF-hose fitting DN 4/6
for SS316 sample gas in/out 4/6mm pipe stubs
Condensate outlet
without condensate pump
1/4′ NPTi or 3/8* NPTi
Protection classIP 20 (EN 60529)
Approvals / SignsCE

Electrics

Power supply220 to 240 VAC 50/60 Hz or 100 to 115 VAC 50/60 Hz
Power consumption
(depending on load and ambient temperature)
190 VA
Connection power / Status signalIEC plug
Fusing6,3 AT in IEC plug
On time100 %
Temperature indicationdigital display
Status threshold<0/>+10°C
Status delay05s
Status relayvolt free contact max. 230 VAC / 6 A min. 5 VADC / 5 mA
Connection terminals / Clamping rangespring type terminals 0,5 mm? to ,5 mm?

DOWNLOADS