Smart environmental monitoring solutions

Talk to our sales team. Call +61 2 9622 6327

Request a quote
Back to products

JES-301E1

Gas Sampling Probe

  • Representative continuous gas sampling in processes containing dust and humidity
  • Installation in Ex-zone 1, 2 or 21, 22
  • Extractive gas analysis
  • Emission (CEMS) and process monitoring
  • Removal of dust from the sample gas

Have questions? Call +61 2 9622 6327

Benefits

  • No condensation or chemical reactions of sample gas
  • Convenient filter replacement
  • Low maintenance
  • Safe unattended operation
  • Easy start-up
  • Even for high dust concentrations
  • Reliable protection of the analyzer system

Features

  • Homogeneous heating without cold spots
  • Large active filter surface
  • Filtration from outside to inside
  • Filter element change without tools
  • Different material of filter elements
  • Self limiting PTC-heater
  • Weather protection housing
  • Various options and upgrades
  • Wide range of accessories
  • Representative continuous gas sampling in processes containing dust and humidity
  • Installation in Ex-zone 1, 2 or 21, 22
  • Extractive gas analysis
  • Emission (CEMS) and process monitoring
  • Removal of dust from the sample gas

Model

JES-301E

Description

Stationary explosion proof gas sampling probe for Ex zone 1, 2, 21 or 22

Operation

Temperature

max. 120°C

Operating pressure

max. 2 bara (max. 7 bara optional)

Flow rate

bis zu 600 Ni/h, depending on application

Mounting position

5° to 15° incline (recommended)

Heating-up time

approx. 180 min
Operating range0°C to 55 °C
Ambient temperature-10°C to 55 °C
Mounting positiontorsion angle max. 45°
Flow rate (back purging)approx. 4 Nm°/h at 4 bar,
5 Nm’/h at 6 bar

Construction

Dimensions over all ( W x H x D )

168 x 415 x 316 mm
models with low temperature contact: 174 x393 x313 mm

Dimensions connection box ( W x H x D )

115 x 115 x 64 mm
models with low temperature contact: 85 x 85 x 55 mm (2 pcs.)

Filter element

2 μm ceramic or 0,2 μm ceramic coated

Active filter surface

170 cm2

Mounting flange

DN 65, PN6; form A according to EN 1092-1; SS316L
or 2 ANSI 150 lbs; according to ASME B16.5

Sample gas inlet

G3/4“ female thread

Sample gas outlet

1/8“NPT female thread

Calibration gas connection (option)

pipe 6 mm, SS316

Back purge connection (option)

pipe 6 mm, SS316

Gas wetted materials

SS316L, ceramic, Viton®

Heating elementPTC, self limiting
Protection classheating: IP68
probe: IP43
connection box: IP66/67
Type of protectionheating: @ Il 2 G Ex d IIC T3 / © 11 2 D Ex tD A21 1P65 T200 °C
connection box: @ Il 2 G Ex e I T6/T5 / ® I 2 D Ex tD A21 IP66 T80 °C/T95 °C
connection box 1 low temperature contact: ® I1 2 G Ex eb IC T6/T5 Gb / @ I1 2 D Ex tb IlIC T80°C/T95 °C Db
connection box 2 low temperature contact: ® I| 2 G Ex ia ib lIC T6 Gb / @ I| 2 D Ex to IlIC T80°C Db
Weightca. 11 kg
Weather protection housingSS304, thermal insulated, colour stainless steel natural
Approvals / signsheating: PTB 02 ATEX 1116X : IECEx PTB 07.0055X
connection box: PTB 99 ATEX 3103 : LCIE 02 ATEX 6240 : IECEx PTB 06.0026 / CE

Electrics

Power supply

115 to 230 VAC 50/60 Hz

Power consumption

approx. 80 W

Temperature controller / sensor

approx. 2 A

Electrical connection

100 %

Switching capacity status contact (volt free)

Cable gland M20 x 1,5 / screw type terminals, clamping range 0,2 to 4 mm²
Low temperature contact (option)Intrinsically safe, NT = 75 °C (NO at AT), max. 60 VAC, 1A, min. 5 VDC / 1 mA
cable gland M16 x 15
Fusingexternal, on installation site

DOWNLOADS