Smart environmental monitoring solutions

Talk to our sales team. Call +61 2 9622 6327

Request a quote
Back to products

LNI ZA FID Air C

Zero Air generator with integrated compressor

The AirGen series ZA FID-C Air is able to remove the total hydrocarbons to less than 0.05 ppm. Very compact, the ZA FID-C is available with a low noise built-in pistons compressor, 100% oil-free.

The filtration of the hydrocarbons is performed using a catalytic oven filled with palladium. The reduction of hydrocarbons, including methane, and carbon monoxide decrease the background noise level and gives much better baseline stability, considerably increasing detector sensitivity and ensuring precise analytical results. High efficiency coalescing pre-filtration and post filtration of particulate size greater than 0.01 microns are used to prevent any kind of particles from entering the analytical device. The ZA FID-C Air has a very simple and efficient interface. It can also be managed by the HydroGen series HG PRO and HG BASIC and is ideal to supply Hydrocarbon free Air to FID detectors.

Categories:

Have questions? Call +61 2 9622 6327

  • Easy use and quick installation
  • High flow rate up to 6 L/min
  • Pressure up to 7 bar
  • RS 485 and RS232 as standard
  • Purity: < 0.05 ppm THC
  • Dew-point <-30°C
  • Reduced foot print
  • Stackable with HG BASIC and HG PRO
  • GC-FID
  • GC-NPD
  • GC-FPD
  • GC-PFPD
  • THA
  • Gas sensing
Models: ZA FID Air 1.5 3.0 6.015,030,0
Zero Air outlet
Flow rate (Max)1.5 l/min3.0 l/min6.0 l/m15.0 l/min30.0 l/min
Outlet pressure (Max)7 bar (101.5 psi)
Total hydrocarbon content< 0.05ppm
Air inlet
Air Inlet requirementISO 8573-1:2010 Class [2:4:1]
Max inlet hydrocarbon content100ppm
Min supply pressure3 bar (43psi)
Max supply pressure10 bar (145 psi)
Temperature (min; Max)1°C (34°F); 35°C (95°F)
Communication
RS485Standard
General data
Supply voltage100-240Vac (±10%) 50/60Hz
Connection typeIEC320-C14
Nominal power (Max)70 W (200 W)140 W (240 W)
Fuse rating (5x20mm)4A (250VAC – T)
Dimensions (W x D x H)27 x 47 x 14 cm
Net weight< 8 kg< 10 kg< 12 kg
Connections
Outlet port1/8″ female BSPP
Inlet port1/8″ female BSPP
Water purgeThreaded ISO M5 female
Operating conditions
Temperature5-35°C (41-95°F)
Humidity (max, non condensing)80.00% [5-35°C (41-95°F)]
Noise< 35dB(A)
IP ratingIP20

DOWNLOADS