Infra-Red Gas Filter Correlation allows the MIR 9000 to offer up to ten gas measurements simultaneously, including Nitric Oxide (NO), Nitrogen Dioxide (NO2), Sulphur Dioxide (SO2), Carbon Monoxide (CO), even Hydrogen Chloride (HCl) and Hydrogen Fluoride (HF).
Complete systems would normally comprise of:
-
- Sample extraction and conditioning probe (with integrated temperature, pressure and flow measurement)
- Sample lines
- Automatic calibration units
- Instrument air drying system
- Data acquisition & management system.
Specifications
MEASURABLE PARAMETERS |
Pollutant |
Range
(ppb) |
Product
code |
HCI |
0-15 / 5000 |
– |
HF |
0-20 / 300 |
– |
NO |
0-100 / 5000 |
– |
NO2 |
0-100 / 1000 |
– |
N2O |
0-20 / 1000 |
– |
NOx |
0-200 / 5000 |
– |
SO2 |
0-75 / 5000 |
– |
CO |
0-75 / 1000 |
– |
CH4 |
0-10 / 1000 |
– |
THC |
0-50 / 5000 |
– |
CO2(%)d> |
0-10 / 100 |
– |
O2(%) |
0-10 / 25 |
– |
Lowest / Highest available ranges expressed in mg/m3 (may vary with your site conditions to be indicated on the Site Survey Form)
(*) Min/Max based on 100/200 Dilution Rate (other ranges & dilution rates available upon request)
TECHNICAL SPECIFICATIONS |
QAL1 to EN 15267-3 certified |
YES |
US EPA compliance |
YES |
Sampling technology |
Cold/Dry Extraction |
Sample transfert |
non heated polytube |
back flush function |
YES |
span injection |
YES on analyser or at sample point |
Built-In O2 option |
YES |
Serial link communication |
RS232 / RS422 |
Other Communication |
Ethernet |
Analogue OUTPUTS |
Optional |
Analogue INPUTS |
Optional |
Dimensions (mm) |
• Tight box version: 200 x 400 x 600 (D x W x H)