Fidas® 200 – EN 16450 certified measurement technology

Fidas® 200

The Fidas® 200 fine dust monitor uses the recognized measuring technique of optical light scattering according to ISO 21501-1 on the single particle and is equipped with an LED light source of high light intensity, high light stability, and long service life. The instrument’s calibration can be easily and quickly checked and, if necessary, adjusted at any time, even when installed, using a monodisperse test aerosol.

The sampling system of the Fidas® 200 operates with a volume flow of approx. 0.3 m3/h.

It is equipped with a Sigma-2 sampling head according to VDI 2119, which enables representative sampling even in strong winds, as well as a drying section, which demonstrably prevents falsification of the measurement due to condensation effects at high humidity.


Categories:

Have questions? Call +61 2 9622 6327

FEATURES & BENEFITS

  • Type-approved and certified according to latest EN requirements (EN 15267)
  • Continuous and simultaneous real-time measurement of multiple PM values
  • Additional information on particle number concentration and particle size distribution
  • Long service life
  • Low maintenance
  • External check of calibration on site possible
  • Intuitive and easy to operate
  • Reliable function, very high data availability (> 99 %)
  • Permanent monitoring of status, among others online monitoring of calibration
  • No radioactive material and no consumables
  • Low energy consumption
  • Regulatory pollution control in monitoring networks
  • Ambient air monitoring campaigns
  • Long-term studies
  • Emission source attribution
  • Emission dispersion studies (e.g. fires, volcanoes)
Measuring principle Optical light scattering at single particles
Reported data PM1, PM2.5, PM4, PM10, TSP, CN, particle size distribution, ambient pressure, ambient temperature, rel. ambient humidity
Measurement range (number CN) 0–20,000 particles/cm³
Measurement range (size) 0.18–18 µm (certified range, other measuring ranges on request)
Measurement range (mass) 0–10,000 µg/m³
Measurement uncertainty 9.7% for PM2.5, 7.5% for PM10 (expanded measurement uncertainty according to EN 16450, TÜV Report)
Volume flow 4.8 l/min ≙ 0.3 m³/h ± 3% (24h), compliant with EN 16450
Size channels 64 (32/decade)
Time resolution 1 s–24 h
Interfaces USB, Ethernet (LAN), RS-232, Wi-Fi
User interface Touchscreen, 800 × 480 pixel, 7″ (17.78 cm)
Protocols UIDEP, UDP, ASCII, MODBUS, Bayern-Hessen
Data logger storage Capacity for 2 years continuous operation at 60 s storage interval
Software PDAnalyze
Data acquisition Digital, 20 MHz processor, 256 raw data channels
Light source Long term stable LED
Housing (Gehäuse) Table housing, optional: with mounting brackets for rack-mounting
Operating system Windows 10 IoT Enterprise
Power supply 115–230 V, 50/60 Hz
Installation conditions +5–+40 °C
Response time < 2 s
Sampling head (Probenahmekopf) Passive collector Sigma-2
Dimensions 450 × 320 × 180.5 mm (H × W × D), 19″
Weight Control unit: 9.3 kg, sample head: 2.25 kg, sample tube: 4.5 kg
Sampling system (Probenahmesystem) Drying of the aerosol by IADS (Intelligent Aerosol Drying System)
Noise emission < 70 dB(A)
Fuse T2A
Resolution 0.1 µg/m³
Power consumption Normal operation: 60 W, max. 200 W
Data management Prepared for connection to the Palas Cloud MyAtmosphere (“MyAtmosphere-ready”); Internet access and separate registration required. The MyAtmosphere terms of use apply.

National and international standards

  • VDI 4202-1
  • VDI 4203-3
  • EN 12341
  • EN 14907
  • EN 16450
  • EU-Äquivalenzleitfaden
  • EN 15267-1/-2
  • ISO 21501-1