Environmental Monitoring Solutions
The Fidas® 200 S version is a 19″ plug-in unit mounted in a splash-proof stainless steel control cabinet for outdoor use (temperature range -20 – 50 °C). A larger, air-conditioned control cabinet is available on request, allowing the installation of additional devices. Variants of the Fidas® 200 S are the basic Fidas® 200 and the Fidas® 200 E with a remote sensor (for easier integration into stations with existing roof feed-through).
Categories:
Have questions? Call +61 2 9622 6327
| Measuring principle | Optical light scattering at single particles |
| Reported data | PM1, PM2.5, PM4, PM10, TSP, CN, particle size distribution, ambient pressure, ambient temperature, rel. ambient humidity |
| Measurement range (number CN) | 0–20,000 particles/cm³ |
| Measurement range (size) | 0.18–18 µm (certified range, other measuring ranges on request) |
| Measurement range (mass) | 0–10,000 µg/m³ |
| Measurement uncertainty | 9.7% for PM2.5, 7.5% for PM10 (expanded measurement uncertainty according to EN 16450, TÜV Report) |
| Volume flow | 4.8 l/min ≙ 0.3 m³/h ± 3% (24h), compliant with EN 16450 |
| Size channels | 64 (32/decade) |
| Time resolution | 1 s–24 h |
| Interfaces | USB, Ethernet (LAN), RS-232, Wi-Fi |
| User interface | Touchscreen, 800 × 480 pixel, 7″ (17.78 cm) |
| Protocols | UIDEP, UDP, ASCII, MODBUS, Bayern-Hessen |
| Data logger storage | Capacity for 2 years continuous operation at 60 s storage interval |
| Software | PDAnalyze |
| Data acquisition | Digital, 20 MHz processor, 256 raw data channels |
| Light source | Long term stable LED |
| Housing (Gehäuse) | Table housing, optional: with mounting brackets for rack-mounting |
| Operating system | Windows 10 IoT Enterprise |
| Power supply | 115–230 V, 50/60 Hz |
| Installation conditions | +5–+40 °C |
| Response time | < 2 s |
| Sampling head (Probenahmekopf) | Passive collector Sigma-2 |
| Dimensions | 450 × 320 × 180.5 mm (H × W × D), 19″ |
| Weight | Control unit: 9.3 kg, sample head: 2.25 kg, sample tube: 4.5 kg |
| Sampling system (Probenahmesystem) | Drying of the aerosol by IADS (Intelligent Aerosol Drying System) |
| Noise emission | < 70 dB(A) |
| Fuse | T2A |
| Resolution | 0.1 µg/m³ |
| Power consumption | Normal operation: 60 W, max. 200 W |
| Data management | Prepared for connection to the Palas Cloud MyAtmosphere (“MyAtmosphere-ready”); Internet access and separate registration required. The MyAtmosphere terms of use apply. |