Environmental Monitoring Solutions

Talk to our sales team. Call +61 2 9622 6327

Request a quote
Back to products

Fidas® Smart 100 E | Fine dust measuring device for existing roof openings

Fidas Smart 100 E

Compact measuring instrument for the determination of fine dust

The system works on the 90-degree scattered light measurement principle on a single particle, considering signal duration and shape. Technology and algorithms were developed based on the EN 16450-certified Fidas ® 200.

Automatic calibration tracking of the measurement system allows operation for up to two years without recalibration. If necessary, the calibration status can be checked and corrected using a test dust calibrated by Palas.

Palas aerosol spectrometers are thus the only optical fine dust measuring instruments that can be calibrated against a traceable standard by the user at the point of operation.

Fidas® Smart 100 features Ethernet, WLAN, and mobile phone connectivity. All measured values are calculated and recorded directly and, if desired, can be transferred to Palas’ own cloud MyAtmosphere directly for visualization or further processing.

Categories:

Have questions? Call +61 2 9622 6327

FEATURES & BENEFITS

  • Extended inlet for installation in existing measuring containers
  • Technology based on the certified Fidas® 200 series (EN16450 and MCERTS); simultaneous measurement of Cn, PM1, PM2.5, PM4, PM10
  • High accuracy due to advanced algorithms
  • Long-term stable: up to 2 years of operation without calibration possible.
  • On-site calibration with test dust (NIST traceable) is possible
  • Operation with AC or DC power source
  • Long-life blower for sample airflow
  • Regulated aerosol heating to avoid condensation
  • Regulatory environmental monitoring
  • Construction sites
  • Networks with roads, railways, and ports
  • Smart City
  • Occupational safety
  • Industry
Measuring principle Optical light scattering at single particles
Reported data PM1, PM2.5, PM4, PM10, TSP, CN, particle size distribution, ambient pressure, ambient temperature, rel. ambient humidity, CO2, Air Quality Index, source indication (depending on configuration)
Measurement range (number CN) 0 – 20,000 particles/cm³
Measurement range (size) 0.18 – 18 µm (certified range, other measuring ranges on request)
Measurement range (mass) 0 – 20,000 µg/m³
Measurement uncertainty 9.0% for PM2.5, 9.7% for PM10 (expanded measurement uncertainty according to EN 16450, TÜV Report)
Volume flow 1 l/min ≙ 0.06 m³/h
Size channels 64 (32/decade)
Time resolution 1 s – 24 h
Interfaces USB, Ethernet (LAN), Wi-Fi, 4G (optional via LTE stick)
User interface Touchscreen, 800 × 480 pixel, 5″ (12.7 cm)
Protocols UDP, ASCII, Modbus
Data logger storage 10 GB
Software PDAnalyze
Data acquisition Digital, 22 MHz processor, 256 raw data channels
Light source Long term stable LED
Housing (Gehäuse) Polymer housing with weather protection and tripod/wall/pole mount option
Operating system Windows 10 IoT Enterprise
Power supply 115 – 230 V, 50/60 Hz
Power consumption Normal operation: 15 W, max. 60 W
Installation conditions -20 – +50 °C (weatherproof)
Response time < 2 s
Dimensions 240 × 320 × 190 mm (H × W × D)
Weight 3.9 kg
Sampling system (Probenahmesystem) Drying of the aerosol by compact IADS (Intelligent Aerosol Drying System) – Version E: extended inlet for installation in measuring containers
Noise emission < 40 dB(A)
Resolution 0.1 µg/m³
Data management Prepared for connection to the Palas Cloud MyAtmosphere (“MyAtmosphere-ready”); Internet access and separate registration required. The MyAtmosphere terms of use apply.

National and international standards

  • ISO 21501-1

DOWNLOADS

You have successfully subscribed to our newsletter!