| Model | JCM-300 |
| Performance | standard |
| Cooling principle | Peltier cooling |
| Number of gas paths | 1 |
| Number of Peltier elements | 1 |
| Integrated condensate pump | option |
| Integrated condensate monitoring | option |
| Operation | |
| Gas flow per gas path* | max. 250 Nl/hr |
| Gas inlet temperature* | max. 140 °C; SS heat exchanger: max. 180 °C |
| Gas inlet dew point* | max. 80°C |
| Gas outlet dew point | 5 °C (factory default); adjustable from 0,5 °C to 7,5 °C |
| Dew point stability (for constant inlet conditions) |
±0,1 K |
| Ambient temperature | 5° to 45 °C |
| Cooling capacity total | max. 15 W |
| Operating pressure with condensate pump | 0,2 to 2,2 bara |
| Max. operating pressure without condensate pump | 4,0 bara; SS heat exchanger: max. 19 bara |
| Ready for operation | < 30 min |
| Pressure drop at max. flow rate | 3 mbar |
| Construction | |
| Dimensions over all (W x H x D) | 244 x 308 x 211 mm |
| Installation | wall mounting |
| Mounting position | approx. 6,5 kg |
| Housing / Colour | stainless steel / natural |
| Gas wetted materials (depending on configuration) | PVDF, SS316Ti, FFKM, Duran glass |
| Dead volume per gas path | 67 ml |
| Connection sample gas and condensate outlet with condensate pump |
PVDF-hose fitting DN 4/6 for SS: pipe 6 mm |
| Condensate outlet without condensate pump | 1/4“ NPTf or 3/8“ NPTf for SS: pipe 12 mm |
| Approvals / Signs | CE |
| Electrics | |
| Power supply | 230 VAC 50/60 Hz +/- 10 % or 115 VAC 50/60 Hz +/- 10 % |
| Power consumption (depending on load and ambient temperature) |
30 to 160 VA |
| Connection power | IEC plug |
| Protection class (in default mounting position) | IP 20 (EN 60529) |
| Fusing | lead fuse T2A in IEC plug |
| On time | 100 % |
| Diagnostic / Operation indicator | 1 x bicolour LED with condensate detector: 2 x bicolour LEDs |
| Status threshold | < 0 / > +10 °C |
| Status delay | 0,5 s |
| Status relay | volt free contact, 230 VAC / 2 A, min. 5 VADC / 5 mA |
| Connection terminals / Clamping range | spring type terminals 0,5 mm2 to 2,5 mm2 |
| Threshold condensate detector | factory setting 12 kΩ adjustable 2 to 30 kΩ |