Environmental Monitoring Solutions
The Particles Plus® 8301-AQM1 and 8301-AQM2 Series Handheld Particle Counter and Environmental Monitor measures 0.3 to 25 µm particles with mass concentration and stores indoor air quality measurements of temperature, relative humidity, CO2, and TVOC in the 8301-AQM2.
This instrument is the most versatile handheld Air Monitor available, with advanced power management and the industry’s first sleep mode, allowing for battery operation of periods that can exceed a month on a single charge.
Categories:
Have questions? Call +61 2 9622 6327
Measures 0.3 μm to 25 μm |
0.1 CFM (2.83 LPM) flow rate |
Large easy-to-use icon-driven color touchscreen display |
Approximates mass concentration in μg/m3 and indicates simultaneous PM values |
User-selectable, adjustable particle channel sizes |
8301-AQM1 measures CO2, temperature and humidity. The 8301-AQM2 includes TVOC. |
Stores up to 45,000 sample records |
Connect using Ethernet, USB or (optional) Wireless 802.11 b/g, RS485 or RS232 |
Static or dynamic IP address (DHCP) connects to a local network or the Internet |
Seamless integration into a facility monitoring system with MODBUS RTU, ASCII or TCP |
Internet of Things JSON output allows for cloud based data storage and retrieval |
Included software permits remote operations, data management, diagnostics and more |
Displays and externally prints information with optional printer |
Internal audible alarm with user selectable thresholds for all environmental parameters |
Advanced power management for long battery operation and patented sleep mode |
Easy configuration and transferable from instrument to instrument |
User-friendly field calibration with single or dual point offsets for all sensors |
Remote diagnostics allows for remote service investigation through the Internet |
Lightweight high-impact injection molded plastic enclosure |
Long-life laser diode technology |
2-year limited warranty, extended warranties available. |
Models | 8306 |
Size Range | 0.3 to 25 μm |
Size Channels | Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm variable binning |
Flow Rate | 0.1 CFM (2.83 LPM) / +/- 3% |
Concentration Limit | 15,000,000 Particles/ft³ @ 10% coincidence loss |
Battery Run Time | >10 hours continuous operation |
Light Resource | Long life laser diode |
Counting Efficiency | 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS |
Zero Count | <1 count / 60 minutes (<1 particles / 6 ft³) |
Count Modes | Real-Time Meter and graph, cumulative/differential count/m³and count/ft³, and mass concentration (PM) |
Count Alarms | 1 to 9,999,999 counts |
Calibration | NIST traceable |
Display | 4.3″ (10.9 cm) WQVGA (480×272) color touch screen |
Printer | (Optional) External thermal printer |
Vacuum Source | Internal pump with automatic flow control |
Filtered Exhaust | Internal HEPA filter |
Number of Channels | 6 |
Custom-sized channels | 6 Calibration for custom size channels available |
Audible Alarm | Adjustable built-in alarm |
Battery | Removable Li-ion |
Battery Recharge Time | 4 hours within instrument (<2 hours with external battery charger) |
Reports | ISO 14644-1, EU GMP Annex 1, FS 209E |
Recipes | 50 user-configurable recipes |
Communication Modes | Ethernet and USB. (Optional) Wireless 802.11 b/g, RS485 or RS232. |
Environmental Sensor | Temp and RH probe 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC),15-90% ±2% relative humidity |
Alarm | Alarms on counts for all particle sizes, low battery, sensor failure, environmental sensors and flow |
Standards | ISO 21501-4 and JIS B9921 |
Instrument Calibration | Recommended minimum once per year |
External Surface | High impact injection molded plastic |
Dimensions (L x W x H) | 5.12” x 4.25” x 12.26” (13.0 cm x 10.8 cm x 31.1 cm) includes handle and does not include probes |
Weight | 2.2 lb (1.0 kg) |
Accessories | Operating manual and IMS software on USB flash drive, isokinetic probe, temperature relative humidity sensor, purge filter, battery, data download software, USB cable, power supply and cable |
Optional Accessories | Printed manual, carrying case, spare battery, external battery charger, external printer and isokinetic probes, barbed fittings, IMS-RT monitoring system |
Buffer Memory | 45,000 sample records (rotating buffer) including particle count data, environmental data, locations, annotations, and times. Scrollable on screen or printout. |
Sample Locations | Up to 1,000 locations 20 characters long |
Sample Time | 1 second to 99 hours |
Power | 110 to 240 VAC 50/60 Hz universal in-line power supply |
Operating Conditions | 41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing |
Storage Conditions | 32º to 122ºF (0º to 50ºC) / Up to 98% non-condensing |
Warranty | 2 year limited warranty. Extended warranties available. |