PEM Hydrogen Generator

The HG KUBE is able to generate up to 260 cc/min of Hydrogen with purity better than 99.9996% and pressure up to 9 bar (130 psi). The HG KUBE includes a cell with polymeric membrane (PEM) using distilled water. No acid or alkaline solutions are involved.

The innovative static drying system ensures the maximum grade of hydrogen purity, protects the GC thanks to it’s unique humidity safe guard, it is completely maintenance-free and allows continuous operation, 24 hours a day.

The on-board CPU automatically checks for internal leaks and constantly control the operating parameters to guarantee full safety. Up to 20 units can be connected in parallel mode using an external control box. The touch-screen LCD interface provides simple and user-friendly management of all functions on the unit.

Categories:

Have questions? Call +61 2 9622 6327

PRODUCT FEATURES

  • PEM cell technology
  • Flow rate up to 260 cc/min
  • Pressure up to 9 bar (130psi)
  • RS-232 and RS 485 ports standard
  • Purity: 99.9996%
  • Static dryer no maintenance
  • Safety routine and self monitoring
  • Reduced foot print
  • Smart humidity safe guard (patented)
  • High pressure GLS electronically controlled (patented)
  • Gas Sensing
  • GC-FID
  • GC-NPD
  • GC-FPD
  • GC-PFPD
  • THA
Model: HG KUBE120260
General data
Electrolytic cellPEM technology
Drying systemStatic dryer
H2 purity>99.9996%
Outlet pressure9 bars (130 psi)
H2 flow-rate (Max)120 cc/min260 cc/min
Dimensions (W x D x H)25 x 30 x 35 cm
Net weight (no water in tank)13 kg
Communication
RS232Standard
RS485Standard
Software functions
Parallel modeOptional
Water
Quality RequiredDeionized, ASTM II, <0.1uS
Supply pressure (min, Max)0.1 bar (1.45 psi), 0.5 bar (7.25 psi)
Supply flow-rate (min, Max)0.2 l/min, 1.5 l/min
Internal tank capacity1.2 L
Electrical data
Type of connectionIEC 320-C14 (desktop adapter)
Power supply voltage100-240 Vac (±10%) 50/60 Hz
Nominal power (Max)160W
Connections
Hydrogen outlet1/8” compression fitting
WaterQuick release push-in fitting
Operating/storage conditions
Working Temperature5-40°C (41-104°F)*1
Storage Temperature1-50°C (34-122°F)
Humidity (max, non condensing)70%
Noise< 35dB(A)
IP ratingIP20
Pollution degree rating2 (with no aromatic compounds)
Altitude< 2000m
*1 referred to O2, dew point <-20°C (< -4°F), Atmospheric Dew Point (ADP)
*2 with temperature > 35°C (> 95°F) the H2 purity could be less