Smart environmental monitoring solutions

Talk to our sales team. Call +61 2 9622 6327

Request a quote
Back to products

HG ST Basic

PEM Hydrogen Generator

The HydroGen series HG ST BASIC is able to generate up to 600 cc/min of Hydrogen with purity better than 99.9999% and pressure up to 10 bar (145 psi). The HG ST BASIC includes a new long life multilayer electrolytic cell with polymeric membrane (PEM) using distilled water. No acid or alkaline solutions are used. The innovative drying system ensures the maximum grade of hydrogen purity, protects the GC thanks to its innovative humidity safe guard, HG ST BASIC is completely maintenance-free and allows continuous operation, 24 hours a day.

The on-board CPU automatically check for internal leaks and constantly control the operating parameters to guarantee full safety. Up to 20 units can be connected in parallel mode using an external control box.The touch-screen LCD interface provides simple and user-friendly management of all functions on the unit.

Categories:

Have questions? Call +61 2 9622 6327

  • Quick installation
  • Easy use
  • Hydrogen flow rate up to 600cc/min
  • Pressure up to 10 bar (145 psi)
  • Hydrogen purity: 99.9999%
  • RS 485 and USB as standard
  • LAN optional
  • Dryer with no maintenance
  • Reduced foot print
  • High pressure GLS electronically controlled (patented)
  • GC-FID / GC-NPD / GC-FPD
  • THA
Models: HG ST BASIC 120 160 250 350 600
General data
Electrolytic cell PEM technology
Drying system Static dryer
H2 purity >99.9999% 1
Outlet pressure 10 bar (145 psi)
H2 flow-rate cc/min (Max) 120 160 250 350 600
Dimensions (WxHxD) 9.5 x 44 x 48 cm
Net weight (no water in tank) 19 kg
Communication
RS485/USB Standard
LAN Optional
Software functions
Parallel mode Ready on board
Automatic tank refill Standard
Water
Quality Deionized, ASTM II, <0.1uS
Supply pressure (min, Max) 0 bars (0 psi), 0.5 bar (7.25 psi)
Internal tank capacity 0.3 l
Electrical data
Type of connection IEC 320-C14
Power supply voltage 100-240 Vac (%±10%) 50/60 Hz
Nominal power (Max) 350W 500W
Fuse rating (5x 20 mm) 4A (250VAC – T)
Connections
Hydrogen outlet 1/8” compression fitting
Water (inlet, drain) Quick release push-in fitting
¹Referred to O2 – DewPoint < -25°C (< -13°F), Atmospheric Dew Point (ADP)

DOWNLOADS

You have successfully subscribed to our newsletter!