Environmental Monitoring Solutions
The HydroGen series HGA RACK 19” 2U BASIC is able to generate up to 600 cc/min of Hydrogen with purity better than 99.9999% and pressure up to 10 bar (145 psi). An AirGen output can deliver hydrocarbon free pure Air with a flow up to 2000 cc/min. The HGA 2U 19 inch BASIC includes a new long life multilayer electrolytic cell with polymeric membrane (PEM) using distilled water. No acid or alkaline solutions are used.
The innovative drying system ensures the maximum grade of hydrogen purity, protects the GC thanks to its innovative humidity safe guard, HGA 19 inch 2U BASIC is completely maintenance-free and allows continuous operation, 24 hours a day. The on-board CPU automatically checks for internal leaks and constantly controls the operating parameters to guarantee full safety. Up to 20 units can be connected in parallel mode using an external control box. The touch-screen LCD interface provides simple and user-friendly management of all functions on the unit.
Categories:
Have questions? Call +61 2 9622 6327
Models: HGA 2U PRO | 100 | 160 | 250 | 350 | 600 |
General data | |||||
Electrolytic cell | PEM technology | ||||
Drying system | Static dryer | ||||
H2 purity | >99.9999% 1 | ||||
Outlet pressure | 10 bars (145 psi) | ||||
H2 flow-rate cc/min (Max) | 100 | 160 | 250 | 350 | 600 |
Dimensions | Standard 19 inch 2U RACK (48 cm deep) | ||||
Net weight (no water in tank) | 17 kg | ||||
Communication | |||||
RS485/USB | Standard | ||||
LAN | Optional | ||||
Software functions | |||||
Parallel mode | Ready on board | ||||
Automatic tank refill | Standard | ||||
Water | |||||
Quality | Deionized, ASTM II, <0.1uS | ||||
Supply pressure (min, Max) | 0 bars (0 psi), 0.5 bar (7.25 psi) | ||||
Internal tank capacity | 0.3 l | ||||
Electrical data | |||||
Type of connection | IEC 320-C14 | ||||
Power supply voltage | 100-240 Vac (±10%) 50/60 Hz | ||||
Installed power (Max) | 350W | 500W | |||
Fuse rating (5x 20 mm) | 4A (250VAC – T) | ||||
Connections | |||||
Hydrogen outlet | 1/8” compression fitting | ||||
Water (inlet, drain) | Quick release push-in fitting | ||||
ZeroAir section | |||||
Inlet pressure (Max) | 8 bars (116 psi) | ||||
THC (Inlet Air) | 100ppm | ||||
Outlet flow rate (max) | 2000 cc/min | ||||
Outlet pressure (max) | Inlet pressure – 0.5 bars (8 psi) at maximum flow | ||||
THC (ZeroAir Outlet) | <0.1ppm | ||||
Inlet, Outlet ports | 1/8″ BSPP female |