Environmental Monitoring Solutions
Gas monitor for fast, precise, direct and continuous measurement of NO2 concentrations in the range 0-1000 ppb.
Visible (450 nm) absorption measurement using patented Cavity Attenuated Phase Shift (CAPS) technology. QAL1 certified by the TÜV, US-EPA approved.
Fast and direct measurement of the sample – no chemical conversion required, no Toxic Gas Emissions.
Categories:
Have questions? Call +61 2 9622 6327
| Measured parameter | NO₂ |
| Technology | CAPS (Cavity Attenuated Phase Shift Spectroscopy) |
| US-EPA Approval / QAL 1 Certification | YES |
| Measurement ranges | 0-1 ppm |
| Measurement units | ppb, ppm, µg/m3 (user-selectable) |
| Lower Detectable Limit (2σ) | 0.1 ppb |
| Noise (σ) | 0.05 ppb |
| Zero drift | 0.75 ppb/24h |
| Span drift | 1.5 ppb/12h |
| Linearity | < 4% |
| Sample flow rate | 1L/min |
| Sampling pump | Internal |
| Temperature range | +10°C to 35°C |
| Data storage | 1 year |
| Included I/O | 1 LAN / 1 USB / WIFI / 1 RS4i board |
| Weight (Kg) | 12.5 |
| Dimensions mm (LxWxH) | 483 x 606 x 133 |
| Chassis | 19″ Rack, 3U |
| Power supply | ‘100~250Vac, 50/60Hz + ground |
COMPLIANCE WITH: