Model | JCP-300 | JCP-310 | ||||||
Performance of sample gas cooler | standard | high-performance | ||||||
Cooling principles | Peltier Cooling | |||||||
Operation | ||||||||
Gas flow per gas path* | max. 250 Ni/hr | |||||||
Gas inlet temperature* | max. 140 °C for PVDF and max. 180 °C for SS heat exchangers | |||||||
Gas inlet dew point* | max. 80 °C | |||||||
Gas outlet dew point | +5°C (factory default); adjustable from +0,5 °C to +7,5 °C | |||||||
Dew point stability (for constant inlet conditions) | +0,1 K | |||||||
Ambient temperature | +5°C to +40 °C | +5°C to +50 °C | ||||||
Cooling capacity total | max. 30 W | max. 60 W | ||||||
Operating pressure with condensate pump | 0.2 to 2.2 bara | |||||||
Max. operating pressure without condensate pump | 4,0 bara; SS heat exchanger: max. 19 bara | |||||||
Ready for operation | < 15 min | |||||||
Pressure drop at max. flow rate | 3 mbar | |||||||
CONSTRUCTION | ||||||||
Dimensions over all (W x H x D) | 561 x 457 x 262 mm | |||||||
Installation | portable | |||||||
Housing / Colour | hard case with trolley function, polypropylene, yellow | |||||||
Operating position | open case, vertical heat exchanger | |||||||
Weight | approx. 18 kg, depending on configuration | approx. 22 kg, depending on configuration | ||||||
Gas wetted materials (depending on configuration) | aluminum coated, PVDF, SS316Ti, FFKM, Duran glass, PTFE, FPE | |||||||
Dead volume per gas path | 67 ml | |||||||
Connection sample gas inlet | stainless steel compression fitting 6 mm | |||||||
Connection sample gas outlet, condensate outlet | PVDF-hose fitting DN 4/6 | |||||||
Approvals / Signs | CE | |||||||
Electrics | ||||||||
Power supply | 230 VAC 50/60 Hz +/-10 % or 115 VAC 50/60 Hz +/-10 % | |||||||
Power consumption (without sample line) (depending on load and ambient temperature) |
60 VA up to 110 VA | 120 VA up to 320 VA | ||||||
Connection power | plug CEE 7/7 to IEC plug, with 2,0 m cable | |||||||
Protection class | transport in closed case: IP 67 (EN 60529) operation with open case: IP 20 (EN 60529) |
|||||||
Fusing (rating depending on sample line) | built-in lead fuse | built-in lead fuse | ||||||
On time | 100 % | |||||||
Diagnostic / Operation indicator | 2x bicolour-LED | |||||||
Status threshold | <0/> +10°C | |||||||
Status relay delay | 0.5 s | |||||||
Status relay | volt free contact. 230 VAC / 2 A, min. 5 VADC / 5 mA | |||||||
Connection for heated sample line | 7-pole plug, load of max. 2200 W or max. 10 A | |||||||
Threshold condensate detector (option) | factory setting 12 kQ2, adjustable 2 to 30 kQ | |||||||
factory setting 12 kQ2, adjustable 2 to 30 kQ | TC Type K (NiCrNi) or RTD PT100 (IEC751) |