Environmental Monitoring Solutions
MP101M Beta gauge monitor with its optional OPM (Optical Particulate Monitor)
The standard Beta gauge measurement Method ISO 10473 of the MP101M analyzer allows, when used with the optical technology of the OPM module, the continuous and simultaneous measurement of fine dust.
The beta attenuation instrument is compliant with EN 16450 for PM10 and PM2.5 European Standards. QAL1 certified by the TÜV, the MP101M is also approved as Federal Equivalent Method (FEM) by US EPA for PM10 and PM2.5 continuous suspended particulate monitoring.
Categories:
Have questions? Call +61 2 9622 6327
The MP101M, based on the beta attenuation measurement technique, determines the particle concentration by measuring the amount of radiation that a sample, collected on a fiber tape, absorbs when exposed to a radioactive source.
Low-energy beta rays are absorbed by the collision of dust, whose number is proportional to density. Absorption is thus a function of the mass of the irradiated material, independently of its physicochemical nature. The OPM’s (Optical Particulate Monitor) principle is based on the measurement of the light scattered by the particles. A powerful algorithm is applied to continuously convert the signal into mass concentration.
The combination of both technologies provides simultaneously approved particulate measurement PLUS real-time indication of PM10, PM2.5, and PM1.
Measured parameter | PM10, PM2.5, PM1, TSP |
Technology | Beta Ray Attenuation |
US-EPA Approval / QAL 1 Certification | YES |
Measurement ranges | up to 10 000 μg/m3 |
Measurement units | µg/m3 |
Lower Detectable Limit (2σ) | 0.5 μg/m3 (24h average) |
Sample flow rate | 1 m³/h |
Sampling pump | External (vacuum pump) |
Temperature range | +5°C to 40°C |
Data storage | Unlimited |
Included I/O | LAN/USB/WIFI/AO |
Weight (Kg) | 15.2 (+pump: Picolino 4.6Kg / KNF 9.4Kg) |
Dimensions mm (LxWxH) | 360.5 x 483 x 266 |
Chassis | 19″ Rack, 3U |
Pressure and temperature compensation | YES |
Technology | Light scattering (*) |
Max. concentration | 0-1000 μg/m3 |
Range of size | 0.3-10 μm |
Lower detection limit | 1 μg/m3 |
Temporal resolution | 1 second |
Dimensions (WxDxH) | 230 x 370 x 200 mm |
Flow | 2.5 L/min |