Environmental Monitoring Solutions
The NitroGen series CASTORE MP 4000 are Membrane Nitrogen Generators. They are engineered to transform standard compressed air into a safe and regulated supply of Nitrogen with a purity up to 99.9%. An additional channel can provide up to 40 L/min of compressed, dry and regulated air.
The CASTORE MP 4000 with its dual Nitrogen and Air outlets has been developed to meet specific requirements in term of gas, flow, purity and pressure for Agilent MP AES 4xxx series. It can be used for the evaporation of solvents in samples being analysed.
Categories:
Have questions? Call +61 2 9622 6327
Model: NG CASTORE MP 4000 | |
N2 outlet | |
Flow rate (Max) | 30 l/min |
Outlet pressure (min / Max) | 2.0 bars / Inlet pressure – 1.0 bar (14.5 psi) |
Nitrogen purityInlet Pressure >= 8 bar (116 psi) | > 99.9 % @ 25 l/min> 99.7 % @ 30 l/min |
Internal tank capacity | 10 L |
Dry Air outlet | |
Flow rate (Max) | 40 l/min |
Outlet pressure | Inlet pressure – 1.0 bar (14.5 psi) |
Dew point *1 | < -30°C (-22°F) |
Internal tank capacity | 10 L |
Air inlet | |
Air inlet requirement | ISO 8573-1:2010 Class [2:4:1] |
Flow rate (Max) | 170 l/min |
Supply pressure (min / Max) | 7 bars (101 psi) / 10 bars (145 psi) |
Temperature (min / Max) | 1°C (34°F); 45°C(113°F) |
General data | |
Supply rating | 100-240Vac (±10%) 50/60 Hz |
Connection type | IEC320-C14 |
Nominal power (Max) | 25W (30VA) |
Fuse rating (5x20mm) | 1A (250Vac – T) |
Net weight | < 40 kg |
Dimensions (W x D x H) | 48 x 35 x 120 cm |
Connections | |
N2 outlet port | ¼” BSPP female |
Dry Air outlet port | ¼” BSPP female |
Inlet port | ¼” BSPP female |
Drain port | ¼” BSPP female |
Operating/storage conditions | |
Temperature | 5-35°C (41-95°F)*2 |
Humidity (max, non condensing) | 80% [5-35°C (41-95°F)] |
Noise | < 35dB(A) |
IP rating | IP20 |
Pollution degree rating | 2 (with no aromatic compounds) |
Altitude | up to 4000m |